U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Medical Device Recalls
  5. Avanos Medical Inc Recalls Certain BALLARD ACCESS Closed Suction Systems for Neonates/Pediatrics for Risk of Inadequate Ventilation, Other Injuries from Cracked Manifolds
  1. Medical Device Recalls

Avanos Medical Inc Recalls Certain BALLARD ACCESS Closed Suction Systems for Neonates/Pediatrics for Risk of Inadequate Ventilation, Other Injuries from Cracked Manifolds

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.

Recalled Product

Device Use

The manifold of the BALLARD ACCESS Closed Suction System for Neonates/Pediatrics is used to provide access to the artificial airway of a child, infant, or neonate, without breaking the ventilation circuit. It is a single patient use device that is used by trained medical professionals.

The BALLARD ACCESS Closed Suction manifold is indicated for use together with the BALLARD ACCESS Closed Suction catheter to remove secretions from the artificial airway of children, infants, and neonates

Reason for Recall

Avanos Medical Inc. is recalling the Avanos BALLARD ACCESS Closed Suction System for Neonates/Pediatrics with Y-Manifold or Elbow Manifold (labeled 72-hour-use) after receiving complaints from customers about cracked manifolds during use. A cracked manifold can cause leaks in the respiratory circuit, which may lead to the patient receiving inadequate ventilation.

Inadequate ventilation and oxygenation can lead to lead serious brain damage or death, especially in the vulnerable patient population (children, infants, and neonates) who receive care using this product. Additional risks from this issue include the possibility that a foreign body could be introduced through the cracked manifold, causing infection, damage to the airway, or a blocked endotracheal tube.

Use of this product could lead to serious injury or death.

Avanos Medical, Inc. reports four injuries associated with this issue.

Who May be Affected

  • Children, infants, or neonates who receive care using a BALLARD ACCESS Closed Suction Systems.
  • Health care personnel providing care for pediatric or neonatal patients with a BALLARD ACCESS Closed Suction Systems.

What to Do

On February 3, 2023, Avanos Medical Inc. sent an Urgent Medical Device Recall letter to customers. The letter included the following recommended actions:

  • Check all storage and usage locations for impacted BALLARD ACCESS Closed Suction System for Neonates/Pediatrics with Y-Manifold or with Elbow Manifold (labeled “72 hour use”).
  • Quarantine all unused impacted product.
  • Devices currently in use with patients.
  • Check for cracks in manifold of any product currently in use with patients.
  • Replace devices with cracked manifolds with a 24-hour use Ballard Closed Suction System for Neonates/Pediatrics (24-hour use products are unaffected by the reported issue).
  • Use the 24-hour-use Ballard Closed Suction System for Neonates/Pediatrics for any new placements or scheduled placements.
  • Complete and return the Acknowledgement Response Form included with the letter to FieldActioncare@avanos.com, with the subject line: FCA-2023-002 Hospital Acknowledgement Response Form.

Customers will be contacted by Avanos Customer Service to arrange for product return, credit, and replacement after the acknowledgement form is submitted.

Contact Information

Customers with questions about this recall should email Avanos at FieldActionCare@avanos.com.

Additional Resources:

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program using an online form, regular mail, or FAX.

 
Back to Top