U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Medical Device Recalls
  5. Datascope/Maquet/Getinge Recalls Cardiosave Hybrid and Rescue Intra-aortic Balloon Pumps (IABPs) for Unexpected Shutdowns
  1. Medical Device Recalls

Datascope/Maquet/Getinge Recalls Cardiosave Hybrid and Rescue Intra-aortic Balloon Pumps (IABPs) for Unexpected Shutdowns

Update: August 31, 2023

The devices described in this recall are the same devices announced in the UPDATE: Risk of Device Failures for Getinge’s Maquet/Datascope Cardiosave Intra-Aortic Balloon Pump (IABP) – Letter to Health Care Providers on August 31, 2023.

Please be aware, this recall is a voluntary correction, not a product removal.

The FDA updated this recall notice to include the June 5, 2023 Important Medical Device Advisory letter sent to all affected customers from Getinge/Maquet/Datascope. 

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.

Recalled Product

  • Recalled Product:
    • Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) and Cardiosave Rescue Intra-Aortic Balloon Pump (IABP)
  • Product Models:
  • Distribution Dates: March 6, 2012 to May 19, 2023
  • Devices Distributed in the U.S.: 4586
  • Date Initiated by Firm: June 5, 2023

Device Use

The Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) and the Cardiosave Rescue IABP are electromechanical systems used to inflate and deflate intra-aortic balloons. These systems provide temporary support to the left ventricle through counter pulsation. Once the balloon is positioned in the aorta, the pump is set to work in synchrony with the electrocardiogram or arterial pressure waveform to make the balloon inflate and deflate at the right time during the cardiac cycle.

Cardiosave Intra-Aortic Balloon Pumps are indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure in adults. They are used in health care facilities.

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) and the Cardiosave Rescue IABP

Reason for Recall

Datascope/Maquet/Getinge is recalling the Cardiosave Hybrid and Rescue Intra-aortic Balloon Pumps (IABPs) because they may shutdown unexpectedly due to electrical failures in the Power Management Board or Solenoid Board (Power Source Path).

Using an affected pump may cause serious adverse health events, including unstable blood pressure, injury (for example: inadequate blood supply or a vital organ injury), and death.

Datascope/Maquet/Getinge reported 26 complaints. There have been no reports of injuries, or death.

Who May be Affected

  • People who receive circulatory support using a Cardiosave Hybrid or Rescue IABP
  • Health care personnel providing care that includes the Cardiosave Hybrid or Rescue IABP

What to Do

On June 5, 2023, Datascope/Maquet/Getinge sent all affected customers an Important Medical Device Advisory.

The letter requested customers to:

  • Ensure there is an alternative IABP available to continue therapy.
  • Provide alternative hemodynamic support if there is no other means to continue counterpulsation therapy.

Contact Information

Customers with questions about this recall should contact their Datascope/Maquet/Getinge representative or call Datascope/Maquet/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time).

Additional Resources:

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program using an online form, regular mail, or FAX.

 

Subscribe to Medical Device Safety and Recalls

Sign up to receive email updates on medical device recalls, safety communications, and other safety information.

Back to Top