In Vitro Diagnostics EUAs - Antigen Diagnostic Tests for SARS-CoV-2
In Vitro Diagnostic EUAs: Overview and Templates
On This Page:
Individual EUAs for Antigen Diagnostic Tests for SARS-CoV-2
This table includes information about authorized SARS-CoV-2 antigen diagnostic tests that have been authorized individually. These EUAs have been issued for each individual test with certain conditions of authorization required of the manufacturer and authorized laboratories. Test attributes are listed in the "Attributes" column. For example, tests authorized for the screening of asymptomatic individuals without known exposure are listed with "screening" in the attribute column; pooling, multi-analyte, saliva, home collection, and home testing are similarly listed. Tests available without a prescription include the attribute "DTC" (for direct-to-consumer home collection tests) or "OTC" (for over-the-counter at-home tests).
For information on tests that have been revoked see Historical Information about Device Emergency Use Authorizations: In Vitro Diagnostics (IVD)
Tests with "single target" in the attribute column are:
- designed to detect only one antigen target, currently either a section of the spike protein or a section of the nucleocapsid protein;
- more susceptible to changes in performance due to viral mutations, meaning they are more likely to fail to detect new variants.
Tests with "multiple targets" in the attribute column are:
- designed to detect more than one section of the proteins that make up SARS-CoV-2;
- more likely to continue to perform as labeled as new variants emerge.
To see additional authorization documents, such as letters granting EUA amendments or revisions, and a list of other brand names authorized under a specific EUA, select the plus (+) button beside the "Date EUA Issued or Last Updated" for each EUA.
Authorized EUA devices in the table below for SARS-CoV-2 antigen diagnostic tests are assigned the QKP product code. Antigen multi-analyte respiratory panel tests authorized in the table below are assigned the QMN product code.
Date EUA Issued or Last Updated | Entity | Diagnostic (Most Recent Letter of Authorization) and Date EUA Originally Issued | Attributes | Authorized Setting(s)1 | Authorization Documents2 | Other Documents | Other Brand Name(s) |
---|---|---|---|---|---|---|---|
08/12/2022 | PHASE Scientific International, Ltd. | INDICAID COVID-19 Rapid Antigen At-Home Test 03/16/2022 | Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | |
07/13/2022 | SD Biosensor, Inc. | Pilot COVID-19 At-Home Test 12/24/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | |
07/08/2022 | iHealth Labs, Inc. | iHealth COVID-19 Antigen Rapid Test 11/05/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | GoToKnow COVID-19 Antigen Rapid Test | |
06/23/2022 | Celltrion USA, Inc. | Celltrion DiaTrust COVID-19 Ag Home Test 10/21/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Multiple Targets | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | None |
09/02/2022 | ACON Laboratories, Inc | Flowflex COVID-19 Antigen Home Test 10/04/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | On/Go One COVID-19 Antigen Home Test | |
08/24/2022 | ANP Technologies, Inc | NIDS COVID-19 Antigen Rapid Test Kit 09/24/2021 | Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
03/25/2022 | Becton, Dickinson and Company (BD) |
BD Veritor At-Home COVID-19 Test |
Lateral Flow, Digital Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, Individuals, IFU, IFU (Home Test) | None | None |
06/11/2021 | Quidel Corporation | Sofia SARS Antigen FIA 05/08/2020 |
Lateral Flow, Fluorescence, Instrument Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
12/10/2021 | Becton, Dickinson and Company (BD) | BD Veritor System for Rapid Detection of SARS-CoV-2 07/02/2020 |
Chromatographic Digital Immunoassay, Instrument Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
07/21/2022 | LumiraDx UK Ltd. | LumiraDx SARS-CoV-2 Ag Test 08/18/2020 |
Microfluidic Immunofluorescence Assay, Instrument Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
02/04/2022 | Abbott Diagnostics Scarborough, Inc. | BinaxNOW COVID-19 Ag Card 08/26/2020 |
Lateral Flow, Visual Read, Single Target | H, M, W | HCP, Patients, IFU | None | |
10/02/2020 | Quidel Corporation | Sofia 2 Flu + SARS Antigen FIA 10/02/2020 |
Lateral Flow, Fluorescence, Instrument Read, Multi-Analyte, Single Target | H, M, W | HCP, Patients, IFU | None | |
12/02/2021 | Access Bio, Inc. | CareStart COVID-19 Antigen test 10/08/2020 |
Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU |
None |
|
10/23/2020 | Celltrion USA, Inc. | Sampinute COVID-19 Antigen MIA 10/23/2020 |
Magnetic Force-assisted Electrochemical Sandwich Immunoassay (MESIA), Single Target | H, M | HCP, Patients, IFU | None | |
03/04/2022 | Luminostics, Inc. | Clip COVID Rapid Antigen Test 12/07/2020 |
Lateral flow immunoluminescent assay, instrument read, Single Target | H, M, W | HCP, Patients, IFU | None | |
07/13/2022 | Ellume Limited | Ellume COVID-19 Home Test 12/15/2020 |
Lateral Flow, Fluorescence, Instrument Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test), FAQ | None | |
05/06/2022 | Abbott Diagnostics Scarborough, Inc. | BinaxNOW COVID-19 Ag Card Home Test 12/16/2020 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Telehealth Proctor Supervised, Serial Screening, Single Target | Home, H, M, W | HCP, Individuals, IFU, IFU (Home Test) | None | |
11/09/2021 | Quidel Corporation | QuickVue SARS Antigen Test 12/18/2020 |
Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | None |
11/16/2021 | Ortho Clinical Diagnostics, Inc. | VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack 01/11/2021 |
Chemiluminescence Immunoassay, Instrument Read, Single Target | H, M | HCP, Patients, IFU | None | |
10/27/2021 | Princeton BioMeditech Corp. | Status COVID-19/Flu A&B 02/04/2021 |
Lateral Flow, Visual Read, Multi-analyte, Single Target | H, M, W | HCP, Patients, IFU | None | |
03/01/2021 | Quidel Corporation | QuickVue At-Home COVID-19 Test 03/01/2021 |
Lateral Flow, Visual Read, Prescription Home Testing, Single Target | Home, H, M, W | HCP, Patients, IFU, IFU (Home Test) | None | |
03/24/2021 | Becton, Dickinson and Company (BD) | BD Veritor System for Rapid Detection of SARS-CoV-2 & Flu A+B 03/24/2021 |
Chromatographic Digital Immunoassay, Instrument Read, Multi-analyte, Single Target | H, M, W | HCP, Patients, IFU Viral Mutation Revision Letter - September 23, 2021 | None | |
09/23/2022 | Quidel Corporation | QuickVue At-Home OTC COVID-19 Test 03/31/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, Individuals, IFU, IFU (Home Test) | CVS Health At Home COVID-19 Test Kit | |
05/05/2022 | Abbott Diagnostics Scarborough, Inc. | BinaxNOW COVID-19 Antigen Self Test 03/31/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, Individuals, IFU, IFU (Home Test) | None | |
02/04/2022 | Abbott Diagnostics Scarborough, Inc. | BinaxNOW COVID-19 Ag 2 Card 03/31/2021 |
Lateral Flow, Visual Read, Non-prescription Testing, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
02/16/2022 | DiaSorin, Inc. | LIAISON SARS-CoV-2 Ag 03/26/2021 |
CLIA, Single Target | H, M | HCP, Patients, IFU | None | |
07/29/2022 | Qorvo Biotechnologies, LLC. | Omnia SARS-CoV-2 Antigen Test 04/13/2021 |
Bulk Acoustic Wave (BAW) Biosensor, Instrument Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
08/11/2022 | Celltrion USA, Inc. | Celltrion DiaTrust COVID-19 Ag Rapid Test 04/16/2021 |
Lateral Flow, Visual Read, Serial Screening, Multiple Targets | H, M, W | HCP, Patients, IFU | Humasis COVID-19 Ag Test | |
05/26/2022 | InBios International, Inc. | SCoV-2 Ag Detect Rapid Test 05/06/2021 |
Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
06/09/2022 | Salofa Oy | Sienna-Clarity COVID-19 Antigen Rapid Test Cassette 05/20/2021 |
Lateral Flow, Visual Read, Single Target | H, M, W | HCP, Patients, IFU |
|
|
08/03/2022 | OraSure Technologies, Inc. | InteliSwab COVID-19 Rapid Test Rx 06/04/2021 |
Lateral Flow, Visual Read, Prescription Home Testing, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | |
08/03/2022 | OraSure Technologies, Inc. | InteliSwab COVID-19 Rapid Test 06/04/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening , Single Target | Home, H, M, W | HCP, Individuals, IFU, IFU (Home Test) | None | |
08/03/2022 | OraSure Technologies, Inc. | InteliSwab COVID-19 Rapid Test Pro 06/04/2021 |
Lateral Flow, Visual Read, Serial Screening , Single Target | H, M, W | HCP, Patients, IFU | None | |
07/08/2021 | Ellume Limited | ellume.lab COVID Antigen Test 07/08/2021 |
Lateral Flow, Fluorescence, Instrument Read, Single Target | H, M, W | HCP, Patients, IFU | None | None |
08/10/2022 | GenBody Inc. | GenBody COVID-19 Ag 07/13/2021 |
Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
08/12/2022 | PHASE Scientific International, Ltd. | INDICAID COVID-19 Rapid Antigen Test 07/28/2021 |
Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
08/29/2022 | Access Bio, Inc. | CareStart COVID-19 Antigen Home Test 08/02/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | On/Go COVID-19 Antigen Self-Test | |
05/27/2022 | QIAGEN GmbH | QIAreach SARS-CoV-2 Antigen 08/05/2021 |
Digital Lateral Flow, Fluorescence, Instrument Read, Single Target | H, M | HCP, Patients, IFU | None | None |
10/12/2021 | Xtrava Health | SPERA COVID-19 Ag Test 10/12/2021 |
Lateral Flow, Visual Read, Single Target | H, M, W | HCP, Patients, IFU | None | None |
09/28/2022 | InBios International Inc. | SCoV-2 Ag Detect Rapid Self-Test 11/22/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | |
02/01/2022 | Nano-Ditech Corp. | Nano-Check COVID-19 Antigen Test 12/06/2021 |
Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | |
8/5/2022 | Siemens Healthineers | CLINITEST Rapid COVID-19 Antigen Self-Test 12/29/2021 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | |
01/14/2022 | iHealth Labs, Inc. | iHealth COVID-19 Antigen Rapid Test Pro 01/14/2022 |
Lateral Flow, Visual Read, Serial Screening, Single Target | H, M, W | HCP, Patients, IFU | None | None |
07/22/2022 | Maxim Biomedical, Inc. | MaximBio ClearDetect COVID-19 Antigen Home Test 01/19/2022 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | |
05/04/2022 | Oceanit Foundry LLC | ASSURE-100 Rapid COVID-19 Test 02/28/2022 |
Lateral Flow, Visual Read, Single Target | H, M, W | HCP, Patients, IFU | None | |
05/24/2022 | Siemens Healthcare Diagnostics, Inc. | Atellica IM SARS-CoV-2 Antigen (CoV2Ag) 03/11/2022 |
CLIA, Single Target | H, M | HCP, Patients, IFU | None | None |
05/24/2022 | Siemens Healthcare Diagnostics, Inc. | ADVIA Centaur SARS-CoV-2 Antigen (CoV2Ag) 03/11/2022 |
CLIA, Single Target | H, M | HCP, Patients, IFU | None | None |
08/19/2022 | OSANG LLC | OHC COVID-19 Antigen Self Test 04/06/2022 | Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Testing, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) |
QuickFinder COVID-19 Antigen Self Test |
|
09/02/2022 | Xiamen Boson Biotech Co., Ltd.” | Rapid SARS-CoV-2 Antigen Test Card 04/06/2022 | Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Testing, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | |
07/08/2022 | Watmind USA | Speedy Swab Rapid COVID-19 Antigen Self-Test 07/08/2022 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Testing, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None | None |
08/23/2022 | Genabio Diagnostics Inc. | Genabio COVID-19 Rapid Self-Test Kit 07/08/2022 |
Lateral Flow, Visual Read, Over the Counter (OTC) Home Testing, Serial Screening, Single Target | Home, H, M, W | HCP, IFU, IFU (Home Test) | None |
1 Authorized settings include the following:
- H - Laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, that meet requirements to perform high complexity tests.
- M - Laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, that meet requirements to perform moderate complexity tests.
- W - Patient care settings operating under a CLIA Certificate of Waiver.
2 Authorization Documents include the Healthcare Provider (HCP) and Patient Fact Sheets and either the Manufacture Instructions/Package Insert (abbreviated to IFU) or the EUA Summary.
Revision Concerning Viral Mutations
On September 23, 2021, the FDA revised the EUAs of certain molecular, antigen, and serology tests to establish additional Conditions of Authorization in response to the continued emergence of new variants of SARS-CoV-2. The revision requires test developers to update their authorized labeling and evaluate the impact of SARS-CoV-2 viral mutations on their test's performance as outlined in the letter. This revision is effective as of September 23, 2021 for all EUAs that are within the scope of the revision. This revision does not apply to EUAs for authorized IL-6 assays or standalone specimen collection devices and does not apply to EUAs that include substantially equivalent viral mutation conditions of authorization.
The FDA has determined that establishing additional conditions is necessary to mitigate the potential risk of false negative results due to either decreased sensitivity or non-reactivity associated with SARS-CoV-2 viral mutations. As set forth in the September 23, 2021 letter, developers of authorized tests that are within the scope of the revision are now required to routinely monitor emerging viral mutations and their potential impact on the performance of the authorized SARS-CoV-2 test(s). If potential impacts are identified, the EUA holder must communicate with the FDA and end users about the potential risk that presence of the mutations may have on test performance. The EUA holder must also update their authorized labeling consistent with the revision letter and submit the labeling to the FDA within 3 months of September 23, 2021. By taking these steps, the FDA and the test developer can quickly act in response to the potential risks identified and, when applicable, share the findings on SARS-CoV-2 Viral Mutations: Impact on COVID-19 Tests.