U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Device Advice: Comprehensive Regulatory Assistance
  4. Unique Device Identification System (UDI System)
  5. UDI Exceptions, Alternatives, and Time Extensions
  1. Unique Device Identification System (UDI System)

UDI Exceptions, Alternatives, and Time Extensions

The UDI rule provides a method for labelers to request exceptions, alternatives, and time extensions to UDI requirements.

On this page:


UDI Policy Regarding Certain Class I and Unclassified Devices1,2

The FDA has issued a guidance on its policy regarding compliance dates for class I and unclassified devices that are not implantable, life-supporting, or life-sustaining. Figure 1 below references the requirements that the FDA does not intend to enforce before September 24, 2022.

Requirement Regulation
UDI Labeling 21 CFR 801.20 & 801.50
Direct Mark 21 CFR 801.45
GUDID Data Submission 21 CFR 830.300
Standard Date Format 21 CFR 801.18

Figure 1

This guidance also describes the FDA's direct mark compliance policy. The FDA does not intend to enforce the direct mark requirements under 21 CFR 801.45 for finished class III, LS/LS, and class II non-sterile devices, requiring a direct mark, that are manufactured and labeled prior to their applicable direct mark compliance date, and that remain in inventory, as well as for finished class I and unclassified devices that are non-sterile, that are manufactured and labeled prior to September 24, 2022, and that remain in inventory, provided the device bears a non-UDI direct mark and the labeler has developed a method by which, using the non-UDI mark, the UDI may be made available. For more information on the policies that apply to certain devices subject to UDI direct marking, see Figure 2, below. 

Resources for policies that apply to certain devices subject to UDI Direct Marking

Category of device  Additional information pertaining to direct mark requirements
Class I and unclassified devices Non-sterile Manufactured and labeled prior to 9/24/2022 Unique Device Identification: Policy Regarding Compliance Dates for Class I and Unclassified Devices and Certain Devices Requiring Direct Marking.  Immediately in Effect Guidance for Industry and Food and Drug Administration Staff
Class II devices other than implantable1, life-sustaining, and life-supporting (I/LS/LS) devices Non-sterile Manufactured and labeled prior to 09/24/2018 Unique Device Identification: Policy Regarding Compliance Dates for Class I and Unclassified Devices and Certain Devices Requiring Direct Marking.  Immediately in Effect Guidance for Industry and Food and Drug Administration Staff
Sterile Manufactured and labeled prior to 09/24/2016 Alternative UDI-A170001
LS/LS devices Non-sterile Manufactured and labeled prior to 09/24/2015 Unique Device Identification: Policy Regarding Compliance Dates for Class I and Unclassified Devices and Certain Devices Requiring Direct Marking.  Immediately in Effect Guidance for Industry and Food and Drug Administration Staff
Sterile Alternative UDI-A170001
Class III devices Non-sterile Manufactured and labeled prior to 09/24/2016 Unique Device Identification: Policy Regarding Compliance Dates for Class I and Unclassified Devices and Certain Devices Requiring Direct Marking.  Immediately in Effect Guidance for Industry and Food and Drug Administration Staff
Sterile Manufactured and labeled prior to 09/24/2014 Alternative UDI-A170001
Consigned and loaned devices of all classes Non-sterile or sterile Consigned and loaned prior to applicable label compliance date
09/24/2014 for class III
09/24/2015 for LS/LS
09/24/2016 for class II
09/24/2018 for class I
Unique Device Identification: Direct Marking of Devices: Guidance for Industry and Food and Drug Administration Staff

Figure 2

1Implantable devices are not subject to direct mark requirements.


Exceptions and Alternatives

General exceptions from the requirement for the label of a device to bear a unique device identifier (§ 801.30)

Under § 801.30, the UDI rule provides general exceptions from UDI labeling requirements to certain categories of devices. A device within one or more of these exceptions is not required to bear a UDI. A labeler of a device identified in § 801.30 is not required to request an exception from the FDA.

Request for an exception from or alternative to a UDI requirement (§ 801.55)

A labeler may submit a request for an exception from or alternative to the requirement for the label of a device to bear a unique device identifier (§ 801.20) or other UDI requirement under 21 CFR 801 Subpart B (Labeling Requirements for Unique Device Identification) for a specified device or a specified type of device.

In response to labeler requests or on our own initiative, the FDA may grant an exception or alternative if an exception is appropriate because the requirements of 21 CFR 801 Subpart B are not technologically feasible, or that an alternative would provide for more accurate, precise, or rapid device identification than the requirements of 21 CFR 801 Subpart B or would better ensure the safety or effectiveness of the device that would be subject to the alternative. If the FDA grants an exception or alternative, we may include any safeguards or conditions deemed appropriate to ensure the adequate identification of the device through its distribution and use. The FDA is making its decisions on labeler requests for exceptions and alternatives available at "FDA Decisions."

Considerations for Labelers

In order to establish a system to adequately identify medical devices through distribution and use, the FDA expects that the labels of almost all devices are capable of bearing, and should in fact bear a unique device identifier (UDI) on the label, unless excepted under § 801.30.

Under 21 CFR 801.55(c), the FDA will consider granting requests for an exception from a UDI requirement if the requirement is not technologically feasible. We expect that such situations will be rare. The FDA does not consider exception requests that are based on reasons other than technological infeasibility (including, but not limited to, financial burden, a claimed low rate of adverse events, or a claim that the product is somehow unique such that adverse events do not occur).

Labelers may consider the following to address issues of inadequate label size or unique packaging that may create challenges to having the device label bear its UDI in both easily readable plain-text and AIDC forms:

  1. Remove or minimize information on the label that is not required under 21 CFR 801 (or 21 CFR 809.10, if your device is an in vitro diagnostic product), or otherwise required by regulation or order. See 21 CFR 801.15 for more information on prominence of required label statements.
  2. Increase the size of the label or modify the label e.g., move label to a flatter location on the immediate container) to accommodate the UDI.
  3. Use a smaller form of AIDC technology or split the AIDC form into multiple segments. The easily readable plain-text UDI may also be split into multiple segments.

If the approaches described above cannot be used to address a labeler's concerns, the labeler may consider submitting a request for an alternative under 21 CFR 801.55 to add an overwrap that would bear the UDI or place another label bearing the UDI elsewhere on the packaging.

To request an exception from or alternative to the requirements of 21 CFR 801 Subpart B:

  • Submit a UDI exception/alternative inquiry below. In response, the FDA UDI Help Desk will email instructions for requesting an exception from or alternative to a UDI requirement. If you do not receive an immediate reply in your inbox, please check the spam/junk folder. If the email was sent to your spam/junk folder, please adjust your filter to recognize the UDI Help Desk as a contact.
  • Review the instructions and include the necessary information in your request.
  • Submit the request as indicated in the instructions. An FDA UDI Help Desk Analyst will respond to your request.

Help Desk Exceptions/Alternatives Inquiry

Submit New Exception or Alternative Inquiry

According to 21 CFR 801.55(a), when submitting your request for an exception or alternative you must:

  • Identify the device or devices that would be subject to the exception or alternative;
  • Identify the provisions of 21 CFR 801 Subpart B that are the subject of the request for an exception or alternative;
  • If requesting an exception, explain why the requirements of 21 CFR 801 Subpart B are not technologically feasible;
  • If requesting an alternative, describe the alternative and explain why it would provide for more accurate, precise, or rapid device identification than the requirements of 21 CFR 801 Subpart B or how the alternative would better ensure the safety or effectiveness of the device that would be subject to the alternative;
  • If known, provide the number of labelers and the number of devices that would be affected if we grant the requested exception or alternative; and
  • Provide other requested information needed to clarify the scope and effects of the requested exception or alternative.

FDA Decisions

The FDA is making its decisions on labeler requests for exceptions and alternatives available to the public on this web page. According to § 801.55(c), any labeler may make use of an exception or alternative granted under § 801.55, provided that such use satisfies all safeguards or conditions that are part of the FDA exception or alternative decision.

Labelers should refer to the information in the table below, especially in the "FDA Decision" column, to determine if an exception or alternative listed may apply to their situation. Labelers may document the applicability and use of an exception/alternative in the device master record (DMR) as part of the labeling specifications required under § 820.181(d) for each respective device, along with a copy of the applicable exception/alternative decision. 

FDA UDI
Exception or
Alternative Number
Keywords FDA Product Codes Request FDA Decision Decision Category Expiration Date
of Exception
or Alternative
Date of Posting
UDI-A160001 UPC [Product Codes] N/A [Decision] Granted with Conditions 2023-09-24 2021-07-29
UDI-A160002 UPC [Product Codes] N/A [Decision] Granted with Conditions 2023-09-24 2021-07-29
UDI-A170001 Sterile inventory;
3-year exception;
21 CFR 801.30(a)(1)
N/A N/A [Decision] Granted with Conditions 2024-09-24: Class III;
2025-09-24: I/LS/LS;
2026-09-24: Class II
2019-04-29

UDI Time Extensions

The FDA requires device labelers to comply with UDI requirements according to the dates established in conjunction with the UDI Rule. The UDI label and GUDID submission compliance date for class II devices that are not considered implantable, life-supporting, or life-sustaining (I/LS/LS) was September 24, 2016.

The FDA believes the phased implementation schedule outlined in the final UDI Rule has given labelers of class III, I/LS/LS, and class II devices adequate time to comply with all applicable UDI requirements by September 24, 2016. We expect all labelers to be making diligent efforts to fulfill their UDI obligations. Labelers with questions about UDI implementation may contact the FDA UDI Help Desk.

The FDA has issued guidance documents on its policy regarding compliance dates for certain class I and unclassified devices and of the requirements for removal of National Health Related Item Code (NHRIC) or National Drug Code (NDC) numbers from device labels.  In addition, FDA has granted labelers of particular types of devices an extension of certain UDI compliance dates when the FDA has determined that an extension would be in the best interest of the public health. Those extension letters are listed below.

UDI Extension Letters


Labeler Codes

Continued use of FDA labeler code (§ 801.57)

The UDI Rule includes a provision that rescinds any National Health Related Item Code (NHRIC) or National Drug Code (NDC) number assigned to a medical device (21 CFR 801.57). On the date a medical device must bear a UDI on its label, any NHRIC or NDC numbers assigned to the device are rescinded and may no longer be on the device label or package (21 CFR 801.57(a)). If a device is not required to bear a UDI on its label, any NHRIC or NDC number assigned to that device is rescinded as of September 24, 2018, and may no longer be on the device label or on any device package (21 CFR 801.57(b)). The Enforcement Policy Regarding Use of National Health Related Item Code and National Drug Code Numbers on Device Labels and Packages - Guidance for Industry and Food and Drug Administration Staff, issued on May 21, 2021, describes the Agency's enforcement policy regarding the regulatory prohibition against providing NHRIC or NDC numbers on device labels and device packages. This guidance indicates that the FDA does not intend to object to the use of legacy FDA identification numbers on device labels and device packages, for finished devices that are manufactured and labeled prior to September 24, 2023.

New labeler codes are not assigned by the FDA for the purposes of assigning NDCs to non-drug products or for use under a system for the issuance of UDIs. A labeler who was previously assigned an FDA labeler code to facilitate use of NHRIC or NDC numbers on devices may wish to continue to use that labeler code as part of their UDIs if consistent with their FDA-accredited issuing agency's guidelines. The FDA's regulations at 21 CFR 801.57(c) and (d) provide that a labeler may submit a request to the FDA for continued use of a previously assigned FDA labeler code under a system for the issuance of UDIs by September 24, 2014. As described in the Enforcement Policy Regarding Use of National Health Related Item Code and National Drug Code Numbers on Device Labels and Packages - Guidance for Industry and Food and Drug Administration Staff, the FDA does not intend to object if a labeler incorporates a previously assigned FDA labeler code into its UDI where it submits a request by September 24, 2021.

Subscribe to UDI

Sign up to receive email updates on Unique Device Identification (UDI).

Back to Top