2020 Scorecard Statistics
- December 2020
- November 2020
- October 2020
- September 2020
- August 2020
- July 2020
- June 2020
- May 2020
- April 2020
- March 2020
- February 2020
- January 2020
December 2020
Certified facilities, as of October 1, 2020 | 8,673 | ||
Certification statistics, as of December 1, 2020 | |||
Total certified facilities / Total accredited units | 8,670 / 22,386 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,654 / 12,947 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,407 / 9,429 | ||
FY 2020 inspection statistics, as of December 1, 2020 | |||
Facilities inspected | 1,311 | ||
Total units at inspected facilities | 3,304 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | .8% | ||
Level 2 violation | 11.1% | ||
Percent of inspections with no violation | 88.1% | ||
Total annual mammography procedures reported, as of December 1, 20201 | 39,150,835 |
November 2020
Certified facilities, as of October 1, 2020 | 8,673 | ||
Certification statistics, as of November 1, 2020 | |||
Total certified facilities / Total accredited units | 8,670 / 22,253 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,653 / 12,924 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,353 / 9,318 | ||
FY 2020 inspection statistics, as of November 1, 2020 | |||
Facilities inspected | 684 | ||
Total units at inspected facilities | 1,694 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 1% | ||
Level 2 violation | 11.3% | ||
Percent of inspections with no violation | 87.7% | ||
Total annual mammography procedures reported, as of November 1, 20201 | 39,178,814 |
October 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of October1, 2020 | |||
Total certified facilities / Total accredited units | 8,673 / 22,059 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,659 / 12,878 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,280 / 9,170 | ||
FY 2020 inspection statistics, as of October 1, 2020 | |||
Facilities inspected | 6,525 | ||
Total units at inspected facilities | 16,218 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.4% | ||
Level 2 violation | 12% | ||
Percent of inspections with no violation | 87.5% | ||
Total annual mammography procedures reported, as of October 1, 20201 | 39,368,087 |
September 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of September 1, 2020 | |||
Total certified facilities / Total accredited units | 8,667 / 21,932 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,651 / 12,858 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,221 / 9,063 | ||
FY 2020 inspection statistics, as of September 1, 2020 | |||
Facilities inspected | 5,640 | ||
Total units at inspected facilities | 14,061 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.5% | ||
Level 2 violation | 12.3% | ||
Percent of inspections with no violation | 87.2% | ||
Total annual mammography procedures reported, as of September 1, 20201 | 39,499,634 |
August 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of August 1, 2020 | |||
Total certified facilities / Total accredited units | 8,664 / 21,921 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,648 / 12,889 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,195 / 9,020 | ||
FY 2020 inspection statistics, as of August 1, 2020 | |||
Facilities inspected | 4,615 | ||
Total units at inspected facilities | 11,506 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.6% | ||
Level 2 violation | 12.8% | ||
Percent of inspections with no violation | 86.6% | ||
Total annual mammography procedures reported, as of August 1, 20201 | 39,630,144 |
July 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of July 1, 2020 | |||
Total certified facilities / Total accredited units | 8,676 / 21,890 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,656 / 12,919 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,157 / 8,957 | ||
FY 2020 inspection statistics, as of July 1, 2020 | |||
Facilities inspected | 3,726 | ||
Total units at inspected facilities | 9,285 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.5% | ||
Level 2 violation | 14.4% | ||
Percent of inspections with no violation | 85.1% | ||
Total annual mammography procedures reported, as of July 1, 20201 | 39,764,264 |
June 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of June 1, 2020 | |||
Total certified facilities / Total accredited units | 8,672 / 21,825 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,652 / 12,919 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,111 / 8,889 | ||
FY 2020 inspection statistics, as of June 1, 2020 | |||
Facilities inspected | 3,582 | ||
Total units at inspected facilities | 8,894 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.5% | ||
Level 2 violation | 14.4% | ||
Percent of inspections with no violation | 85.1% | ||
Total annual mammography procedures reported, as of June 1, 20201 | 39,796,305 |
May 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of May 1, 2020 | |||
Total certified facilities / Total accredited units | 8,698 / 21,709 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,647 / 12,883 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,062 / 8,809 | ||
FY 2020 inspection statistics, as of May 1, 2020 | |||
Facilities inspected | 3,582 | ||
Total units at inspected facilities | 8,894 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.5% | ||
Level 2 violation | 14.4% | ||
Percent of inspections with no violation | 85.1% | ||
Total annual mammography procedures reported, as of May 1, 20201 | 39,792,833 |
April 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of April 1, 2020 | |||
Total certified facilities / Total accredited units | 8,683 / 21,739 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,663 / 12,942 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 6,041 / 8,779 | ||
FY 2020 inspection statistics, as of April 1, 2020 | |||
Facilities inspected | 3,582 | ||
Total units at inspected facilities | 8,894 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.5% | ||
Level 2 violation | 14.4% | ||
Percent of inspections with no violation | 85.1% | ||
Total annual mammography procedures reported, as of April 1, 20201 | 39,793,446 |
March 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of March 1, 2020 | |||
Total certified facilities / Total accredited units | 8,695 / 21,671 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,676 / 12,953 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 5,989 / 8,697 | ||
FY 2020 inspection statistics, as of March 1, 2020 | |||
Facilities inspected | 3,194 | ||
Total units at inspected facilities | 7,909 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.6% | ||
Level 2 violation | 14.2% | ||
Percent of inspections with no violation | 85.3% | ||
Total annual mammography procedures reported, as of March 1, 20201 | 39,840,776 |
February 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of February 1, 2020 | |||
Total certified facilities / Total accredited units | 8,696 / 21,563 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,671 / 12,951 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 5,928 / 8,588 | ||
FY 2020 inspection statistics, as of February 1, 2020 | |||
Facilities inspected | 2,533 | ||
Total units at inspected facilities | 6,230 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.6% | ||
Level 2 violation | 14.6% | ||
Percent of inspections with no violation | 84.8% | ||
Total annual mammography procedures reported, as of February 1, 20201 | 39,748,060 |
January 2020
Certified facilities, as of October 1, 2019 | 8,668 | ||
Certification statistics, as of January 1, 2020 | |||
Total certified facilities / Total accredited units | 8,690 / 21,421 | ||
Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,663 / 12,952 | ||
Certified facilities with DBT3,4 units / Accredited DBT units | 5,834 / 8,443 | ||
FY 2020 inspection statistics, as of January 1, 2020 | |||
Facilities inspected | 1,953 | ||
Total units at inspected facilities | 4,785 | ||
Percent of inspections where the highest noncompliance was a: | |||
Level 1 violation | 0.5% | ||
Level 2 violation | 14.6% | ||
Percent of inspections with no violation | 84.9% | ||
Total annual mammography procedures reported, as of January 1, 20201 | 39,730,720 |