U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Medical Device Recalls
  5. Abbott Vascular Recalls NC Trek RX and NC Traveler RX Coronary Dilatation Catheters Due to Failure of Balloon (diameter 4.0mm, 4.5mm and 5.00mm) to Deflate
  1. Medical Device Recalls

Abbott Vascular Recalls NC Trek RX and NC Traveler RX Coronary Dilatation Catheters Due to Failure of Balloon (diameter 4.0mm, 4.5mm and 5.00mm) to Deflate

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.

Recalled Product:

  • NC Trek RX Coronary Dilatation Catheter and NC Traveler RX Coronary Dilatation Catheters
    Models: Balloon diameters 4.0mm, 4.5mm and 5.0mm
    Lots: See a full list of affected devices
  • Distribution Dates: August 2019 to January 2020
  • Devices Recalled in the U.S.: 13,891
  • Date Initiated by Firm: January 29, 2020
Image of the Abbott Vascular NC Trek RX Coronary Dilatation Catheter

Image of the Abbott Vascular NC Trek RX Coronary Dilatation Catheter

Device Use:
Coronary dilation catheters are used to open clogged blood vessels to improve blood flow to the heart.

The NC TREK RX Coronary Dilatation Catheter is indicated for use in the following cardiac (heart) procedures:

  • balloon dilatation of the abnormal narrowing portion of a coronary artery or bypass graft, for the purpose of improving myocardial perfusion.
  • balloon dilatation of a coronary artery occlusion, to restore coronary flow in patients who have had a type of heart attack, known as myocardial infarction – specifically patients with ST-segment elevation, which refers to the flat section of an electrocardiogram (ECG) reading and represents the interval between heartbeats.
  • balloon dilatation of a stent after implantation (balloon models 4.00 mm – 5.00 mm only).

The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.

Reason for Recall:
Abbot Vascular is recalling the NC Trek RX Coronary Dilatation Catheter and NC Traveler RX Coronary Dilatation Catheters, balloon diameters 4.0mm, 4.5mm and 5.0mm because the balloons from the impacted lots may not deflate as intended. This issue is due to weaker material close to the balloon bond resulting from excessive exposure to heat during manufacturing.

Use of these devices may cause serious adverse health consequences, such as prolonged cardiac ischemia (reduced blow flow to the heart), air embolism, thrombosis (clot in the artery), myocardial infarction (heart attack), and additional surgery that could lead to post-operative complications, including death.

Abbot Vascular has received 13 complaints related to this issue. One death has been reported.

Who May be Affected

  • Patients who undergo cardiac procedures involving the NC Trek RX Coronary Dilatation Catheter and NC Traveler RX Coronary Dilatation Catheters.
  • Health care providers and hospitals using the NC Trek RX Coronary Dilatation Catheter and NC Traveler RX Coronary Dilatation Catheters.

What to Do
On January 29, 2020, Abbot Vascular issued an Urgent Recall Notification instructing customers to:

  • Immediately stop using affected devices from the lots affected by this recall.
  • Review your inventory, complete, and return the attached Effectiveness Check Form.
  • Return all unused affected product to Abbott Vascular.
  • Share the notification with other relevant personnel in your organization.

Contact Information
Customers who have questions or need additional information about this recall should contact your local Abbott Vascular Representative or Customer Service department at: 800.227.9902.

Full List of Affected Devices

Device Identifier/GTIN Device Description Part Number Lot Numbers
08717648152054 NC TREK RX 4.00 X 8MM BDC 1012453-08 90731G1 90808G1 90826G1 90906G1 90921G1 90928G1 91003G1 91017G1 91106G1
08717648152061 NC TREK RX 4.00 X 12MM BDC 1012453-12 90730G1 90731G1 90818G1 90822G1 90905G1 90918G1 90919G1 90927G1 91001G1 91015G1 91025G1 91101G1
08717648152078 NC TREK RX 4.00 X 15MM BDC 1012453-15 90815G1 90816G1 90828G1 90904G1 90920G1 90926G1 91008G1 91009G1 91020G1 91026G1 91109G1 91117G1
08717648152085 NC TREK RX 4.00 X 20MM BDC 1012453-20 90727G1 90830G1 90923G1 91004G1 91028G1
08717648152092 NC TREK RX 4.50 X 8MM BDC 1012454-08 90801G1 90812G1 90818G1 90818G2 90904G1 90912G1 90925G1 91010G2 91022G1 91025G1 91101G1
08717648152108 NC TREK RX 4.50 X 12MM BDC 1012454-12 90731G1 90805G1 90817G1 90818G1 90904G1 90906G1 90912G1 90922G1 91015G1 91016G1 91105G1
08717648152115 NC TREK RX 4.50 X 15MM BDC 1012454-15 90819G1 90819G2 90916G1 90916G2 90927G1 90930G1 91031G1
08717648152122 NC TREK RX 4.50 X 20MM BDC 1012454-20 90801G1 90809G1 91018G1 91021G1
08717648152139 NC TREK RX 5.00 X 8MM BDC 1012455-08 90918G1 90930G1 91001G1
08717648152146 NC TREK RX 5.00 X 12MM BDC 1012455-12 90918G1 90926G1 90930G1 91031G1
08717648152153 NC TREK RX 5.00 X 15MM BDC 1012455-15 90806G1 90806G2 91022G1 91025G1
08717648152160 NC TREK RX 5.00 X 20MM BDC 1012455-20 91010G1 91026G1
08717648195983 NC Traveler RX 4.0 X 8MM 1013157-08 91010G1
08717648195990 NC Traveler RX 4.0 X 12MM 1013157-12 90812G1
08717648196003 NC Traveler RX 4.0 X 15MM 1013157-15 91102G1
08717648196027 NC Traveler RX 4.5 X 8MM 1013158-08 90812G1
08717648196034 NC Traveler RX 4.5 X 12MM 1013158-12 90813G1

Additional Resources
NC TRAVLER Coronary Dilatation Catheter (4.5 X8MM) - Recall Database Entry
NC TRAVLER Coronary Dilatation Catheter (4.5 X 12MM) - Recall Database Entry
NC TRAVLER Coronary Dilatation Catheter (4.0 X 12MM) – Recall Database Entry
NC TRAVLER Coronary Dilatation Catheter (4.0 X 8MM) – Recall Database Entry
NC TREK Coronary Dilatation Catheter (5.00 X 20MM BDC) – Recall Database Entry
NC TREK Coronary Dilatation Catheter (5.00 X 15MM BDC) – Recall Database Entry
NC TREK Coronary Dilatation Catheter (5.00 X 12MM BDC) – Recall Database Entry
NC TREK Coronary Dilatation Catheter (5.00 X 8MM BDC) – Recall Database Entry
NC TREK Coronary Dilatation Catheter (4.50 X 20MM BDC) - Recall Database Entry
NC TREK Coronary Dilatation Catheter (4.50 X 15MM BDC) – Recall Database Entry

Abbott Vascular Urgent Field Notice

How do I report a problem?
Health care professionals and consumers may report adverse reactions or quality problems they experience using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail or by FAX.

Back to Top