GUIDANCE DOCUMENT
Policy for Coronavirus Disease-2019 Tests During the Public Health Emergency (Revised) Guidance for Developers and Food and Drug Administration Staff November 2021
- Docket Number:
- FDA-2020-D-0987
- Issued by:
-
Guidance Issuing OfficeCenter for Devices and Radiological Health
FDA plays a critical role in protecting the United States from threats such as emerging infectious diseases, including the Coronavirus Disease 2019 (COVID-19) pandemic. FDA is committed to providing timely guidance to support response efforts to this pandemic.
FDA is issuing this guidance to provide FDA's enforcement policies regarding certain novel coronavirus (COVID-19) tests for the duration of the public health emergency. Rapid detection of COVID-19 cases in the United States requires wide availability of testing to control this rapidly spreading, severe illness. This document supersedes "Policy for Coronavirus Disease-2019 Tests During the Public Health Emergency (Revised): Guidance for Clinical Laboratories, Commercial Manufacturers, and Food and Drug Administration Staff" issued May 11, 2020.
The policies in this guidance are intended to remain in effect only for the duration of the public health emergency related to COVID-19 declared by the Secretary of Health and Human Services (HHS) on January 31, 2020, effective January 27, 2020, including any renewals made by the HHS Secretary in accordance with section 319(a)(2) of the Public Health Service Act (PHS Act). FDA continues to assess the evolving situation and intends to update this guidance as appropriate.
Given this public health emergency, and as discussed in the Notice in the Federal Register of March 25, 2020, titled "Process for Making Available Guidance Documents Related to Coronavirus Disease 2019," this guidance is being implemented without prior public comment because FDA has determined that prior public participation for this guidance is not feasible or appropriate (see section 701(h)(1)(C) of the Federal Food, Drug, and Cosmetic Act (FD&C Act) and 21 CFR 10.115(g)(2)). This guidance document is being implemented immediately, but it remains subject to comment in accordance with the Agency's good guidance practices.
Submit Comments
You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))
If unable to submit comments online, please mail written comments to:
Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852
All written comments should be identified with this document's docket number: FDA-2020-D-0987.