GUIDANCE DOCUMENT
Enforcement Policy for Ventilators and Accessories and Other Respiratory Devices During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency Guidance for Industry and Food and Drug Administration Staff March 2020
- Docket Number:
- FDA-2020-D-1138
- Issued by:
-
Guidance Issuing OfficeCenter for Devices and Radiological Health
March 13, 2023 - This guidance is intended to remain in effect until November 7, 2023, unless superseded by a revised final guidance before that date. For further information, refer to 88 FR 15417, March 13, 2023.
March 24, 2023 - The FDA has finalized two guidances: Transition Plan for Medical Devices That Fall Within Enforcement Policies Issued During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency and Transition Plan for Medical Devices Issued Emergency Use Authorizations (EUAs) Related to Coronavirus Disease 2019 (COVID-19). The guidances outline the FDA's general recommendations to transition from certain policies adopted and operations implemented during the COVID-19 pandemic to normal operations, including the FDA's recommendations for:
- Developing a transition implementation plan,
- Submitting a marketing submission, and
- Taking other actions with respect to these devices.
The FDA encourages stakeholders to review the two final guidances, attend the webinar, and reach out to the FDA if they have questions or concerns. In particular, for manufacturers planning to seek marketing authorization for their devices, the FDA recommends beginning work on a marketing submission, including a transition implementation plan, as described in the guidances.
Additional Resources:
- Transition Plan for Medical Devices That Fall Within Enforcement Policies Issued During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency
- Transition Plan for Medical Devices Issued Emergency Use Authorizations (EUAs) Related to Coronavirus Disease 2019 (COVID-19)
- Webinar on Guidances on COVID-19 Transition Plans for Medical Devices - April 18, 2023
The Food and Drug Administration (FDA or Agency) plays a critical role in protecting the United States from threats including emerging infectious diseases, including the Coronavirus Disease 2019 (COVID-19) pandemic. FDA is committed to providing timely guidance to support response efforts to this pandemic.
FDA is issuing this guidance to provide a policy to help expand the availability of ventilators as well as other respiratory devices and their accessories during this pandemic.
This policy is intended to remain in effect only for the duration of the public health emergency related to COVID-19 declared by the Department of Health and Human Services (HHS), including any renewals made by the Secretary in accordance with section 319(a)(2) of the PHS Act.
Given this public health emergency, this guidance is being implemented without prior public comment because the FDA has determined that prior public participation for this guidance is not feasible or appropriate (see section 701(h)(1)(C)(i) of the Federal Food, Drug, and Cosmetic Act (FD&C Act) and 21 CFR 10.115(g)(2)). This guidance document is being implemented immediately, but it remains subject to comment in accordance with the Agency's good guidance practices.
Submit Comments
You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))
If unable to submit comments online, please mail written comments to:
Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852
All written comments should be identified with this document's docket number: FDA-2020-D-1138.