U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Science and Research | Medical Devices
  4. Modified Dot-Blot Method to Evaluate Personal Protective Equipment (PPE)
  1. Science and Research | Medical Devices

Modified Dot-Blot Method to Evaluate Personal Protective Equipment (PPE)

Catalog of Regulatory Science Tools to Help Assess New Medical Devices

Technical Description

The Modified Dot-Blot Method to Evaluate Personal Protective Equipment (PPE) is a peer-reviewed method to test the barrier performance of PPE materials against viral surrogate or liquid (artificial blood) penetration. The physical equipment components needed to execute this method comprise:

  • a Dot-Blot Apparatus (for example, Whatman Schleicher & Schuell Minifold I) modified as described in Appendix A,
  • PPE materials (such as, gowns and drapes) and
  • clinically relevant artificial test soils.

For a visual depiction of the apparatus, refer to Appendix A.

For step-by-step protocols for testing of PPE materials, refer to Appendix B.

Intended Purpose

The intended purpose of the Modified Dot-Blot Method to Evaluate PPE is to serve as a screening method to evaluate the barrier effectiveness of Level 4 PPE (surgical gowns and surgical drapes) against penetration by

  • bacteriophage ΦΧ174 (universal viral surrogate) or
  • clinically relevant test soils.

If the PPE fails preliminary screening by this method, it may eliminate the need for evaluation by FDA-recognized standards ASTM F1671/ F1671M-13 (Standard Test Method For Resistance Of Materials Used In Protective Clothing To Penetration By Blood-Borne Pathogens Using Phi-X174 Bacteriophage Penetration As A Test System) or ASTM F1670/F1670M-17a (Standard Test Method For Resistance Of Materials Used In Protective Clothing To Penetration By Synthetic Blood), which are more resource-intensive.

As will be described in more detail in subsequent sections, the method consists of securing a cut piece of PPE into the modified Dot-Blot apparatus, exposing the PPE to bacteriophage ΦΧ174 or artificial blood for a set time, and applying vacuum for a set time to test the ability of the PPE to resist penetration.

Note: The ASTM methods use a common apparatus that serves as a comparator for the modified Dot-Blot apparatus described here.

Testing

The Modified Dot-Blot Method’s utility in informing the decision whether to evaluate PPE using ASTM F1671/ F1671M-13 is demonstrated in the publication Comparing ASTM F1671 with a Modified Dot-Blot Method to Evaluate Personal Protective Materials. Materials, test organism, and equivalent pressures were identical between the two methods, and the results were essentially equivalent.

The Modified Dot-Blot Method’s utility in informing the decision whether to evaluate PPE using ASTM F1670/F1670M-17a is demonstrated in the publication Evaluation of Apparatus Used to Test Liquid through Protective Materials: Comparison of a Modified Dot-Blot Apparatus to the ASTM Penetration Cell. Pressures, materials, and artificial test soils were identical between the two methods, and the results were essentially equivalent

After evaluating the equivalence of the Modified Dot-Blot Method (as it pertains to artificial blood) to the ASTM F1670 test method, the Modified Dot-Blot Method was used to evaluate several additional clinically relevant test soils representing urine, vomit, and feces. Note: ASTM F1670 does not contain provisions for these test soils. When the Modified Dot-Blot method was used with these test soils, increased failure rates of PPE were observed, as described in the publication Using Clinically Relevant Test Soils to Evaluate Personal Protective Materials. These observations underscore the need for careful consideration of the chemical nature of test soils and their respective effects on PPE.

Limitations

The Modified Dot-Blot Method does not replace or alter the recognition of currently FDA-recognized standards used to evaluate PPE, such as ASTM F1671/ F1671M-13 and ASTM F1670/F1670M-17a. Rather, the Modified Dot-Blot Method can be used to decide whether it is appropriate to undertake these more resource-intensive evaluations. If a PPE device passes screening with the Modified Dot-Blot Method, evaluation using the resource-intensive methods may be necessary.

Limitations of using the Modified Dot-Blot Method as it relates to ASTM F1671 (FDA-recognized standard for evaluating viral penetration of Level 4 gowns) include that the Modified Dot-Blot Method:

  • does not directly detect viable bacteriophage ΦΧ174; it detects antigenic proteins of the bacteriophage. There is a possibility of artifactual detection (false positive).
  • tests a smaller surface area than the ASTM F1671 method. A greater number of tests are needed with the Modified Dot-Blot to assess the same sample surface area.

More details on the difference between F1671 and The Modified Dot-Blot method can be found in the publication Comparing ASTM F1671 with a Modified Dot-Blot Method to Evaluate Personal Protective Materials.

The limitations of using the Modified Dot-Blot Method as it relates to ASTM F1670 (FDA-recognized standard for evaluating liquid penetration of Level 4 drapes) include that the Modified Dot-Blot Method:

  • uses a smaller surface area than the ASTM F1670 method. A greater number of tests are needed with the Modified Dot-Blot to assess the same sample surface area.

More details on the difference between F1670 and The Modified Dot-Blot method can be found in the publication Evaluation of Apparatus Used to Test Liquid through Protective Materials: Comparison of a Modified Dot-Blot Apparatus to the ASTM Penetration Cell

Supporting Documentation

The Materials and Methods sections of the following peer-reviewed publications contain the appropriate level of detail needed to execute the Modified Dot-Blot Method to Evaluate PPE.

The paper Comparing ASTM F1671 with a Modified Dot-Blot Method to Evaluate Personal Protective Materials describes using the Modified Dot-Blot Method with bacteriophage, in direct comparison to the FDA-recognized ASTM standard method.

The paper Evaluation of Apparatus Used to Test Liquid through Protective Materials: Comparison of a Modified Dot-Blot Apparatus to the ASTM Penetration Cell describes using the Modified Dot-Blot Method with blood, in direct comparison to the FDA-recognized ASTM standard method.

The paper Using Clinically Relevant Test Soils to Evaluate Personal Protective Materials describes using the Modified Dot-Blot Method with test soils not covered by any FDA-recognized PPE evaluation standard.

Contact

Tool Reference

In addition to citing relevant publications please reference the use of this tool using DOI: 10.5281/zenodo.7883668.

For more information:


Appendix: Step-by-step intructions

Appendix A

The original unmodified apparatus is currently commercially available. The image depicted in Figure 1 is of the dismantled apparatus after it has been modified. The modification, for the purposes of this method, consists of drilling out the 1 mm holes in the filtration plate (not the sample well plate) to expand the diameter to 5 mm and thus increase the surface area of the PPE sample exposed to the vacuum.

Figure 1 The Modified Dot-Blot Apparatus. The PPE, the nitrocellulose and the blotting paper are placed between the sample well plate and the filtration plate.

Figure 1
The Modified Dot-Blot Apparatus. In this expanded view, the apparatus layers from top to bottom are shown: clamping plate; sample well plate containing a silicone o-ring in each well; a gown sample, nitrocellulose and blotting paper layered together; a filtration plate; and at the bottom a vacuum plenum with the vacuum port showing.

Appendix B. Protocols for Testing PPE Materials

To execute the Modified Dot-Blot Method as it pertains to viral penetration, as a screening test before undertaking the more resource-intensive ASTM F1671/ F1671M-13, the steps are listed below.

  1. Cut PPE, nitrocellulose membrane and blotting paper into rectangles of 12cm x 9cm.
  2. Soak blotting paper in PBS for a minimum of 2 minutes; prewet nitrocellulose membrane in PBS for a minimum of 1 minute.
  3. Place blotting paper, then the nitrocellulose membrane and then PPE on the filtration plate. Be sure the face (outside) of the PPE is directly below the sample wells so it will directly contact test solution. Ensure that it lies flat when the apparatus is re-assembled.
  4. Clamp in place according to apparatus manufacturer’s instructions.
  5. Using a multi-channel pipet, place 100 mL of suspension (test organism in nutrient broth) in wells (any unused wells must be blocked off).
  6. Let rest for 5 minutes.
  7. Apply vacuum of 0.14 bar (2 psi) via the vacuum port for one minute. Turn off vacuum and allow system to return to ambient pressure.
  8. If no visual penetration or loss of test suspension from the wells is observed, continue observing for an additional 54 minutes.
  9. Using a multi-channel pipet or a pipet tip attached to vacuum, remove any remaining test suspension. It is important to do this without dismantling the apparatus to avoid the test suspension contaminating non-leak areas of the nitrocellulose membrane.
  10. Using a multichannel pipet, place 100ml PBS in each well, then remove all liquid from the wells. Repeat this washing step once.
  11. Carefully dismantle the apparatus and use tweezers to remove the nitrocellulose membrane. Place in a shallow dish slightly bigger than the membrane.
  12. Wash the nitrocellulose membrane with PBS-T twice to remove any unbound assay interferents.
  13. Into the dish containing the nitrocellulose membrane, pour a solution of 0.05% BSA (bovine serum albumin) in PBS-Tween 20 and cover the nitrocellulose membrane. Rock gently for 30 minutes. The BSA blocks the unbound sites on the nitrocellulose membrane.
  14. Remove solution from shallow dish, keeping the nitrocellulose membrane.
  15. Rinse the nitrocellulose membrane three times with PBS-Tween 20, rocking gently for 10 min each time. Remove solution after each rinse step, keeping the nitrocellulose membrane.
  16. Add solution of primary antibodies* (in PBS-Tween 20) to bind to viral surrogate. Rock gently for 30 min. Remove solution from shallow dish, keeping the nitrocellulose membrane.
  17. Rinse the nitrocellulose membrane three times with PBS-Tween 20, rocking gently for 10 min each time. Remove solution after each rinse step, keeping the nitrocellulose membrane.
  18. Add solution of secondary antibody* (in PBS-Tween 20) to bind to the primary antibody. Rock gently for 30 min. Remove solution from shallow dish, keeping the nitrocellulose membrane.
  19. Rinse twice with PBS-Tween 20, rocking gently for 10 min each time. Remove solution after each rinse step, keeping the nitrocellulose membrane.
  20. Rinse once with PBS, rocking gently for 10 min. Remove PBS solution, keeping the nitrocellulose membrane.
  21. Add solution containing substrate* specific to the secondary antibody, covering the nitrocellulose membrane. The reacted substrate will change color and precipitate if the secondary antibody is present (i.e., bound to the primary antibody, in turn bound to viral surrogate, in turn bound to the nitrocellulose membrane). Color change is a qualitative indication of PPE failure.
  22. Any evidence of penetration constitutes a “Fail” endpoint.

*The primary antibodies bind to the antigenic proteins of the bacteriophage. Primary antibodies specific to ΦΧ174 are needed. If primary antibodies are not commercially available, contract labs can produce them. Rabbits are commonly used to produce primary antibodies. Primary antibodies, while they bind to the antigenic proteins, generally do not have labels discernible by detection techniques. Such is the case with the present method. Therefore, secondary antibodies with labels are needed to bind to the primary antibodies. An appropriate secondary antibody for this method would be a goat antirabbit with HRP (horse radish peroxidase) enzyme label; it is available from commercial vendors. After the binding of the labeled secondary antibody to the primary antibody, the label is detected through the use of a commercially available substrate for the HRP enzyme. Once the substrate reacts with the enzyme, a signal occurs. In the case of the present method, the reacted substrate changes color and precipitates, allowing for visual detection. The signal-to-noise ratio is modulated by the respective concentrations and incubation times of primary and secondary antibodies; acceptable concentrations for desired signal-to-noise ratios should be determined before using the present method.

To execute the Modified Dot-Blot Method as it pertains to artificial blood, as a screening test before undertaking the more resource-intensive ASTM F1670, the steps are listed below.

  1. Cut PPE and blotting paper into rectangles of 12cm x 9cm.
  2. Soak blotting paper in PBS for a minimum of 2 minutes.
  3. Place blotting paper and then PPE on the filtration plate. Be sure the face (outside) of the PPE is directly below the sample wells so it will directly contact test solution. Ensure that it lies flat when the apparatus is re-assembled.
  4. Clamp in place according to specific manufacturer’s instructions.
  5. Using a multi-channel pipet, place 100 mL of test solution (artificial blood) in wells (any unused wells must be blocked off).
  6. Let rest for 5 minutes.
  7. Apply vacuum of 0.14 bar (2 psi) via the vacuum port for one minute. Turn off vacuum and allow system to return to ambient pressure.
  8. If no visual penetration or loss of test suspension from the wells is observed, continue observing for an additional 54 minutes.
  9. Using a multi-channel pipet or a pipet tip attached to vacuum, remove any remaining test suspension. It is important to do this without dismantling the apparatus to avoid the test solution contaminating non-leak areas of the blotting paper.
  10. Carefully dismantle the apparatus and check if any dye penetrated the PPE and is visible on the blotting paper.
  11. Any evidence of penetration constitutes “Fail” endpoint.
 
Back to Top