Table of Pharmacogenetic Associations
Pharmacogenetic tests, along with other information about patients and their disease or condition, can play an important role in drug therapy. When a health care provider is considering prescribing a drug, knowledge of a patient's genotype may be used to aid in determining a therapeutic strategy, determining an appropriate dosage, or assessing the likelihood of benefit or toxicity.
On this page:
- About the Table
- Section 1: Pharmacogenetic Associations for which the Data Support Therapeutic Management Recommendations
- Section 2: Pharmacogenetic Associations for which the Data Indicate a Potential Impact on Safety or Response
- Section 3: Pharmacogenetic Associations for which the Data Demonstrate a Potential Impact on Pharmacokinetic Properties Only
- Updates to the Table
- Additional Resources
About the Table
For the pharmacogenetic associations listed in this table, the FDA has evaluated and believes there is sufficient scientific evidence to suggest that subgroups of patients with certain genetic variants, or genetic variant-inferred phenotypes (such as affected subgroup in the table below), are likely to have altered drug metabolism, and in certain cases, differential therapeutic effects, including differences in risks of adverse events.
The fact that the FDA has included a particular gene-drug interaction in the table does not necessarily mean the FDA advocates using a pharmacogenetic test before prescribing the corresponding medication, unless the test is a companion diagnostic. Tests that are essential for the safe and effective use of a therapeutic product, including those that identify patients for which the drug is contraindicated, are companion diagnostics.
This table is not intended to affect current regulatory requirements or policies, including the FDA's policy regarding companion diagnostics: Guidance for Industry and FDA Staff: In Vitro Companion Diagnostic Devices. Nor is the table intended to make an assessment on the safe and effective use of, or regulatory requirements for, tests that detect variants in the referenced genes, or to provide comprehensive information on the described gene-drug interactions.
Specific information regarding therapeutic management is provided for some pharmacogenetic associations listed in the table, but most of the associations listed have not been evaluated in terms of the impact of genetic testing on clinical outcomes, such as improved therapeutic effectiveness or increased risk of specific adverse events. In addition, clinical studies, if available, may only have linked genetic variation to a drug's pharmacokinetics (such as the way in which the drug is metabolized), and differences in drug efficacy or safety across different genotype subgroups may not be known. If no statements related to efficacy or toxicity are provided, the scientific evidence the FDA reviewed was considered insufficient to support such associations.
The FDA recognizes that practitioners will take into account different sources and strengths of evidence and will make prescribing decisions based on their judgment about which treatments are appropriate for individual patients. In particular, each patient's genetic makeup is only one of many factors that may impact drug concentrations and response, highlighting the fact that information provided in this table is limited to certain pharmacogenetic associations only and does not provide comprehensive information needed for safe and effective use of a drug. Accordingly, health care providers should refer to FDA-approved labeling for prescribing information, including monitoring instructions and information on other factors that may affect drug concentrations, benefits, and risks. In this context, the information in this Table is intended primarily for prescribers, and patients should not adjust their medications without consulting their prescriber.
This version of the table is limited to pharmacogenetic associations that are related to drug metabolizing enzyme gene variants, drug transporter gene variants, and gene variants that have been related to a predisposition for certain adverse events. The FDA recognizes that various other pharmacogenetic associations exist that are not listed here, and this table will be updated periodically with additional pharmacogenetic associations supported by sufficient scientific evidence.
The FDA has opened a docket for stakeholders—including scientific and medical communities, patients, providers, and industry—to offer specific comments on pharmacogenetic associations that the FDA should or should not include in this table, along with the rationale and underlying evidence that supports the pharmacogenetic association.
Please submit all comments to the open docket. Please submit any questions to [email protected].
Section 1: Pharmacogenetic Associations for which the Data Support Therapeutic Management Recommendations
Drug | Gene | Affected Subgroups+ | Description of Gene-Drug Interaction |
---|---|---|---|
Abacavir | HLA-B | *57:01 allele positive | Results in higher adverse reaction risk (hypersensitivity reactions). Do not use abacavir in patients positive for HLA-B*57:01. |
Amifampridine | NAT2 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. Use lowest recommended starting dosage and monitor for adverse reactions. Refer to FDA labeling for specific dosing recommendations. |
Amifampridine Phosphate | NAT2 | poor metabolizers | Results in higher systemic concentrations. Use lowest recommended starting dosage (15 mg/day) and monitor for adverse reactions. |
Amphetamine | CYP2D6 | poor metabolizers | May affect systemic concentrations and adverse reaction risk. Consider lower starting dosage or use alternative agent. |
Aripiprazole | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. Dosage adjustment is recommended. Refer to FDA labeling for specific dosing recommendations. |
Aripiprazole Lauroxil | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Dosage adjustment is recommended. Refer to FDA labeling for specific dosing recommendations. |
Atomoxetine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. Adjust titration interval and increase dosage if tolerated. Refer to FDA labeling for specific dosing recommendations. |
Azathioprine | TPMT and/or NUDT15 | intermediate or poor metabolizers | Alters systemic active metabolite concentration and dosage requirements. Results in higher adverse reaction risk (myelosuppression). Consider alternative therapy in poor metabolizers. Dosage reduction is recommended in intermediate metabolizers for NUDT15 or TPMT. Intermediate metabolizers for both genes may require more substantial dosage reductions. Refer to FDA labeling for specific dosing recommendations. |
Belinostat | UGT1A1 | *28/*28 (poor metabolizers) | May result in higher systemic concentrations and higher adverse reaction risk. Reduce starting dose to 750 mg/m2 in poor metabolizers. |
Brexpiprazole | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Dosage adjustment is recommended. Refer to FDA labeling for specific dosing recommendations. |
Brivaracetam | CYP2C19 | intermediate or poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. Consider dosage reductions in poor metabolizers. |
Capecitabine | DPYD | intermediate or poor metabolizers | Results in higher adverse reaction risk (severe, life-threatening, or fatal toxicities). No dosage has proven safe in poor metabolizers, and insufficient data are available to recommend a dosage in intermediate metabolizers. Withhold or discontinue in the presence of early-onset or unusually severe toxicity. |
Carbamazepine | HLA-B | *15:02 allele positive | Results in higher adverse reaction risk (severe skin reactions). Avoid use unless potential benefits outweigh risks and consider risks of alternative therapies. Patients positive for HLA-B*15:02 may be at increased risk of severe skin reactions with other drugs that are associated with a risk of Stevens Johnson Syndrome/Toxic Epidermal necrolysis (SJS/TEN). Genotyping is not a substitute for clinical vigilance. |
Celecoxib | CYP2C9 | poor metabolizers or *3 carriers | Results in higher systemic concentrations. Reduce starting dose to half of the lowest recommended dose in poor metabolizers. Consider alternative therapy in poor metabolizers with juvenile rheumatoid arthritis. |
Citalopram | CYP2C19 | poor metabolizers | Results in higher systemic concentrations and adverse reaction risk (QT prolongation). The maximum recommended dose is 20 mg. |
Clobazam | CYP2C19 | intermediate or poor metabolizers | Results in higher systemic active metabolite concentrations. Poor metabolism results in higher adverse reaction risk. Dosage adjustment is recommended. Refer to FDA labeling for specific dosing recommendations. |
Clopidogrel | CYP2C19 | intermediate or poor metabolizers | Results in lower systemic active metabolite concentrations, lower antiplatelet response, and may result in higher cardiovascular risk. Consider use of another platelet P2Y12 inhibitor. |
Clozapine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Dosage reductions may be necessary. |
Codeine | CYP2D6 | ultrarapid metabolizers | Results in higher systemic active metabolite concentrations and higher adverse reaction risk (life-threatening respiratory depression and death). Codeine is contraindicated in children under 12 years of age. |
Deutetrabenazine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and adverse reaction risk (QT prolongation). The maximum recommended dosage should not exceed 36 mg (maximum single dose of 18 mg). |
Dronabinol | CYP2C9 | intermediate or poor metabolizers | May result in higher systemic concentrations and higher adverse reaction risk. Monitor for adverse reactions. |
Eliglustat | CYP2D6 | ultrarapid, normal, intermediate, or poor metabolizers | Alters systemic concentrations, effectiveness, and adverse reaction risk (QT prolongation). Indicated for normal, intermediate, and poor metabolizer patients. Ultrarapid metabolizers may not achieve adequate concentrations to achieve a therapeutic effect. The recommended dosages are based on CYP2D6 metabolizer status. Coadministration with strong CYP3A inhibitors is contraindicated in intermediate and poor CYP2D6 metabolizers. Refer to FDA labeling for specific dosing recommendations. |
Erdafitinib | CYP2C9 | *3/*3 (poor metabolizers) | May result in higher systemic concentrations and higher adverse reaction risk. Monitor for adverse reactions. |
Flibanserin | CYP2C19 | poor metabolizers | May result in higher systemic concentrations and higher adverse reaction risk. Monitor patients for adverse reactions. |
Flurbiprofen | CYP2C9 | poor metabolizers or *3 carriers | Results in higher systemic concentrations. Use a reduced dosage in poor metabolizers. |
Fluorouracil | DPYD | intermediate or poor metabolizer | Results in higher adverse reaction risk (severe, life-threatening, or fatal toxicities). No dosage has proven safe in poor metabolizers and insufficient data are available to recommend a dosage in intermediate metabolizers. Withhold or discontinue in the presence of early-onset or unusually severe toxicity. |
Fosphenytoin | CYP2C9 | intermediate or poor metabolizers | May result in higher systemic concentrations and higher adverse reaction risk (central nervous system toxicity). Consider starting at the lower end of the dosage range and monitor serum concentrations. Refer to FDA labeling for specific dosing recommendations. Carriers of CYP2C9*3 alleles may be at increased risk of severe cutaneous adverse reactions. Consider avoiding fosphenytoin as an alternative to carbamazepine in patients who are CYP2C9*3 carriers. Genotyping is not a substitute for clinical vigilance and patient management. |
Fosphenytoin | HLA-B | *15:02 allele positive | May result in higher adverse reaction risk (severe cutaneous reactions). Patients positive for HLA-B*15:02 may be at increased risk of Stevens Johnson Syndrome/Toxic Epidermal necrolysis (SJS/TEN). Consider avoiding fosphenytoin as an alternative to carbamazepine in patients who are positive for HLA-B*15:02. Genotyping is not a substitute for clinical vigilance and patient management. |
Gefitinib | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. Monitor for adverse reactions. |
Iloperidone | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (QT prolongation). Reduce dosage by 50%. |
Irinotecan | UGT1A1 | *28/*28 (poor metabolizers) | Results in higher systemic active metabolite concentrations and higher adverse reaction risk (severe neutropenia). Consider reducing the starting dosage by one level and modify the dosage based on individual patient tolerance. |
Lofexidine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. Monitor for orthostatic hypotension and bradycardia. |
Meclizine | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May affect systemic concentrations. Monitor for adverse reactions and clinical effect. |
Meloxicam | CYP2C9 | poor metabolizers or *3 carriers | Results in higher systemic concentrations. Consider dose reductions in poor metabolizers. Monitor patients for adverse reactions. |
Metoclopramide | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. The recommended dosage is lower. Refer to FDA labeling for specific dosing recommendations. |
Mercaptopurine | TPMT and/or NUDT15 | intermediate or poor metabolizers | Alters systemic active metabolite concentration and dosage requirements. Results in higher adverse reaction risk (myelosuppression). Initial dosages should be reduced in poor metabolizers; poor metabolizers generally tolerate 10% or less of the recommended dosage. Intermediate metabolizers may require dosage reductions based on tolerability. Intermediate metabolizers for both genes may require more substantial dosage reductions. Refer to FDA labeling for specific dosing recommendations. |
Mivacurium | BCHE | intermediate or poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (prolonged neuromuscular blockade). Avoid use in poor metabolizers. |
Oliceridine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (respiratory depression and sedation). May require less frequent dosing. |
Pantoprazole | CYP2C19 | poor metabolizers | Results in higher systemic concentrations. Consider dosage reduction in children who are poor metabolizers. No dosage adjustment is needed for adult patients who are poor metabolizers. |
Phenytoin | CYP2C9 | intermediate or poor metabolizers | May result in higher systemic concentrations and higher adverse reaction risk (central nervous system toxicity). Refer to FDA labeling for specific dosing recommendations. Carriers of CYP2C9*3 alleles may be at increased risk of severe cutaneous adverse reactions. Consider avoiding phenytoin as an alternative to carbamazepine in patients who are CYP2C9*3 carriers. Genotyping is not a substitute for clinical vigilance and patient management. |
Phenytoin | HLA-B | *15:02 allele positive | May result in higher adverse reaction risk (severe cutaneous reactions). Patients positive for HLA-B*15:02 may be at increased risk of Stevens Johnson Syndrome/Toxic Epidermal necrolysis (SJS/TEN). Consider avoiding phenytoin as an alternative to carbamazepine in patients who are positive for HLA-B*15:02. Genotyping is not a substitute for clinical vigilance and patient management. |
Pimozide | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Dosages should not exceed 0.05 mg/kg in children or 4 mg/day in adults who are poor metabolizers and dosages should not be increased earlier than 14 days. |
Piroxicam | CYP2C9 | intermediate or poor metabolizers | Results in higher systemic concentrations. Consider reducing dosage in poor metabolizers. |
Pitolisant | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Use lowest recommended starting dosage. Refer to FDA labeling for specific dosing recommendations. |
Propafenone | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (arrhythmia). Avoid use in poor metabolizers taking a CYP3A4 inhibitor. |
Sacituzumab Govitecan-hziy | UGT1A1 | *28/*28 (poor metabolizers) | May result in higher systemic concentrations and adverse reaction risk (neutropenia). Monitor for adverse reactions and tolerance to treatment. |
Siponimod | CYP2C9 | intermediate or poor metabolizers | Results in higher systemic concentrations. Adjust dosage based on genotype. Do not use in patients with CYP2C9 *3/*3 genotype. Refer to FDA labeling for specific dosing recommendations. |
Succinylcholine | BCHE | intermediate or poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (prolonged neuromuscular blockade). Avoid use in poor metabolizers. May administer test dose to assess sensitivity and administer cautiously via slow infusion. |
Tacrolimus | CYP3A5 | intermediate or normal metabolizers | Results in lower systemic concentrations, lower probability of achieving target concentrations and may result in higher rejection risk. Measure drug concentrations and adjust dosage based on trough whole blood tacrolimus concentrations. |
Tetrabenazine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. The maximum recommended single dose is 25 mg and should not exceed 50 mg/day. |
Thioguanine | TPMT and/or NUDT15 | intermediate or poor metabolizers | Alters systemic active metabolite concentration and dosage requirements. Results in higher adverse reaction risk (myelosuppression). Initial dosages should be reduced in poor metabolizers; poor metabolizers generally tolerate 10% or less of the recommended dosage. Intermediate metabolizers may require dosage reductions based on tolerability. Intermediate metabolizers for both genes may require more substantial dosage reductions. Refer to FDA labeling for specific dosing recommendations. |
Thioridazine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (QT prolongation). Predicted effect based on experience with CYP2D6 inhibitors. Contraindicated in poor metabolizers. |
Tramadol | CYP2D6 | Ultrarapid metabolizers3 | Results in higher systemic and breast milk active metabolite concentrations, which may result in respiratory depression and death. Contraindicated in children under 12 and in adolescents following tonsillectomy/adenoidectomy. Breastfeeding is not recommended during treatment. |
Valbenazine | CYP2D6 | poor metabolizers | Results in higher systemic active metabolite concentrations and higher adverse reaction risk (QT prolongation). Dosage reductions may be necessary. |
Venlafaxine | CYP2D6 | poor metabolizers | Alters systemic parent drug and metabolite concentrations. Consider dosage reductions. |
Vortioxetine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. The maximum recommended dose is 10 mg. |
Warfarin | CYP2C9 | intermediate or poor metabolizers | Alters systemic concentrations and dosage requirements. Select initial dosage, taking into account clinical and genetic factors. Monitor and adjust dosages based on INR. |
Warfarin | CYP4F2 | V433M variant carriers | May affect dosage requirements. Monitor and adjust doses based on INR. |
Warfarin | VKORC1 | -1639G>A variant carriers | Alters dosage requirements. Select initial dosage, taking into account clinical and genetic factors. Monitor and adjust dosages based on INR. |
Section 2: Pharmacogenetic Associations for which the Data Indicate a Potential Impact on Safety or Response
Drug | Gene | Affected Subgroups+ | Description of Gene-Drug Interaction |
---|---|---|---|
Allopurinol | HLA-B | *58:01 allele positive | Results in higher adverse reaction risk (severe skin reactions). |
Carbamazepine | HLA-A | *31:01 allele positive | Results in higher adverse reaction risk (severe skin reactions). Consider risk and benefit of carbamazepine use in patients positive for HLA-A*31:01. Genotyping is not a substitute for clinical vigilance. |
Carvedilol | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (dizziness). |
Cevimeline | CYP2D6 | poor metabolizers | May result in higher adverse reaction risk. Use with caution. |
Codeine | CYP2D6 | poor metabolizers | Results in lower systemic active metabolite concentrations and may result in reduced efficacy. |
Efavirenz | CYP2B6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (QT prolongation). |
Isoniazid | Nonspecific (NAT) | poor metabolizers | May result in higher systemic concentrations and adverse reaction risk. |
Lapatinib | HLA-DRB1 | *07:01 allele positive | Results in higher adverse reaction risk (hepatotoxicity). Monitor liver function tests regardless of genotype. |
Lapatinib | HLA-DQA1 | *02:01 allele positive | Results in higher adverse reaction risk (hepatotoxicity). Monitor liver function tests regardless of genotype. |
Nilotinib | UGT1A1 | *28/*28 (poor metabolizers) | Results in higher adverse reaction risk (hyperbilirubinemia). |
Oxcarbazepine | HLA-B | *15:02 allele positive | Results in higher adverse reaction risk (severe skin reactions). Patients positive for HLA-B*15:02 may be at increased risk of severe skin reactions with other drugs that are associated with a risk of Stevens Johnson Syndrome/Toxic Epidermal necrolysis (SJS/TEN). Genotyping is not a substitute for clinical vigilance. |
Pazopanib | HLA-B | *57:01 allele positive | May result in higher adverse reaction risk (liver enzyme elevations). Monitor liver function tests regardless of genotype. |
Pazopanib | UGT1A1 | *28/*28 (poor metabolizers) | Results in higher adverse reaction risk (hyperbilirubinemia). |
Perphenazine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk. |
Procainamide | Nonspecific (NAT) | poor metabolizers | Alters systemic parent drug and metabolite concentrations. May result in higher adverse reaction risk. |
Simvastatin | SLCO1B1 | 521 TC or 521 CC (intermediate or poor function transporters) | Results in higher systemic concentrations and higher adverse reaction risk (myopathy). The risk of adverse reaction (myopathy) is higher for patients on 80 mg than for those on lower doses. |
Sulfamethoxazole and Trimethoprim | Nonspecific (NAT) | poor metabolizers | May result in higher adverse reaction risk. |
Sulfasalazine | Nonspecific (NAT) | poor metabolizers | Results in higher systemic metabolite concentrations and higher adverse reaction risk. |
Tolterodine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations and higher adverse reaction risk (QT prolongation). |
Voriconazole | CYP2C19 | intermediate or poor metabolizers | Results in higher systemic concentrations and may result in higher adverse reaction risk. |
Section 3: Pharmacogenetic Associations for which the Data Demonstrate a Potential Impact on Pharmacokinetic Properties Only
The impact of these genetic variants or genetic variant inferred phenotypes on the safety or response of the corresponding drug has not been established.
Drug | Gene | Affected Subgroups+ | Description of Gene-Drug Interaction |
---|---|---|---|
Amitriptyline | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Amoxapine | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Atorvastatin | SLCO1B1 | 521 CC (poor function transporters) | May result in higher systemic concentrations. |
Avatrombopag | CYP2C9 | intermediate or poor metabolizers | Results in higher systemic concentrations. |
Carisoprodol | CYP2C19 | poor metabolizers | Results in higher systemic concentrations. Use with caution. |
Clomipramine | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Darifenacin | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. |
Desipramine | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Dexlansoprazole | CYP2C19 | intermediate or poor metabolizers | Results in higher systemic concentrations. |
Diazepam | CYP2C19 | poor metabolizers | May affect systemic concentrations. |
Dolutegravir | UGT1A1 | poor metabolizers | Results in higher systemic concentrations. |
Donepezil | CYP2D6 | ultrarapid or poor metabolizers | Alters systemic concentrations. |
Doxepin | CYP2C19 | intermediate or poor metabolizers | Results in higher systemic concentrations. |
Doxepin | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Elagolix | SLCO1B1 | 521 CC (poor function transporters) | Results in higher systemic concentrations. |
Escitalopram | CYP2C19 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Esomeprazole | CYP2C19 | poor metabolizers | Results in higher systemic concentrations. |
Fesoterodine | CYP2D6 | poor metabolizers | Results in higher systemic active metabolite concentrations. |
Fluvoxamine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Use with caution. |
Galantamine | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Titrate dosage based on tolerability. |
Hydralazine | Nonspecific (NAT) | poor metabolizers | Results in higher systemic concentrations. |
Ibuprofen | CYP2C9 | poor metabolizers or *3 carriers | May result in higher systemic concentrations. |
Imipramine | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Metoprolol | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. |
Mirabegron | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. |
Nebivolol | CYP2D6 | poor metabolizers | May result in higher systemic concentrations. |
Nortriptyline | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Omeprazole | CYP2C19 | intermediate or poor metabolizers | Results in higher systemic concentrations. |
Paroxetine | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Propranolol | CYP2D6 | poor metabolizers | May affect systemic concentrations. |
Protriptyline | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. |
Rabeprazole | CYP2C19 | poor metabolizers | Results in higher systemic concentrations. |
Raltegravir | UGT1A1 | *28/*28 (poor metabolizers) | Results in higher systemic concentrations. |
Risperidone | CYP2D6 | poor metabolizers | Alters systemic parent drug and metabolite concentrations. |
Rosuvastatin | SLCO1B1 | 521 CC (poor function transporters) | Results in higher systemic concentrations. |
Tamoxifen | CYP2D6 | intermediate or poor metabolizers | Results in lower systemic active metabolite concentrations. The impact of CYP2D6 intermediate or poor metabolism on efficacy is not well established. |
Tamsulosin | CYP2D6 | poor metabolizers | Results in higher systemic concentrations. Predicted effect based on experience with CYP2D6 inhibitors. Use with caution. |
Trimipramine | CYP2D6 | ultrarapid, intermediate, or poor metabolizers | May alter systemic concentrations. |
Viloxazine | CYP2D6 | poor metabolizers | May result in higher systemic concentrations. |
+ The table describes gene-drug interactions and indicates specific affected subgroup(s) to which the interaction applies. The affected subgroup(s) may be carriers of a specific genetic variant (for example, HLA-B*15:02), or a genotype-inferred phenotype, ultrarapid, normal, intermediate, or poor metabolizers/function transporters of a drug metabolizing enzyme/drug transporter. Normal metabolizers or normal transporters do not have genetic variants that are expected to impact metabolism or transport function. In general, ultrarapid metabolizers have two or more copies of a genetic variant that increases metabolic function; intermediate metabolizers or reduced function transporters are individuals who have one or two copies of a genetic variant that reduces the ability to metabolize or transport a drug; and poor metabolizers or poor function transporters are individuals who generally have two copies of a genetic variant that results in little to no ability to metabolize or transport a drug.
When evaluating if a gene-drug association is supported, the FDA believes it is important to ensure that there is sufficient scientific evidence to support:
- The relationship between the detected genetic variant and as appropriate, diplotypes, and the affected subgroup, and
- The association between the affected subgroup and the claimed gene-drug interaction.
In some cases, a specific genetic variant may affect the metabolism of different drugs in different ways. In cases where the association is limited to specific genetic variants and does not apply to all individuals with the genotype-inferred phenotype, the specific variants are provided in the table. In cases where individual genetic variants are not listed in the table, the FDA believes there is sufficient scientific evidence to generally support the described association for the genotype-inferred phenotype subgroup, provided specific genetic variants are determined to confer the genotype-inferred phenotype based on sufficient scientific evidence.
For example, when considering, as described in the table, that poor and intermediate metabolizers of CYP2C19 have higher systemic active metabolite concentrations, higher adverse reaction risk, and dosage adjustments are recommended when taking clobazam, sufficient scientific evidence supports the following, with respect to the *2 allele:
- The functionality of the *2 variant is known, such as *2 variant results in a loss of CYP2C19 enzyme function, and
- *1/*2 confers an intermediate metabolizer phenotype and *2/*2 confers a poor metabolizer phenotype.
Updates to the Table
- November 8, 2021: The FDA made these additions:
- Section 1: Added fosphenytoin and phenytoin.
- Section 3: Added viloxazine.
- May 24, 2021: The FDA made these changes or additions:
- Section 1: Added oliceridine and updated the Affected Subgroups and Description of Gene-Drug Interaction for celecoxib, flurbiprofen, and meloxicam.
- Section 3: Added Ibuprofen.
- March 18, 2021: The FDA made these changes or additions:
- Section 1: Added meloxicam, pitolisant, and sacituzumab. Updated the Description of Gene-Drug Interaction for tacrolimus.
- Section 2: Moved voriconazole from Section 3 to Section 2 and updated the Description of Gene-Drug Interaction.
- Section 3: Added atorvastatin and updated the Affected Subgroups and Description of Gene-Drug Interaction for escitalopram.
- February 25, 2020: The FDA corrected an error for the drug Tramadol. The affected subgroup was previously listed incorrectly as poor metabolizers in the initial version.