U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Medical Device Recalls
  5. Stryker Neurovascular Recalls Trevo XP ProVue Retriever Due to Core Wire That May Break or Separate During Use
  1. Medical Device Recalls

Stryker Neurovascular Recalls Trevo XP ProVue Retriever Due to Core Wire That May Break or Separate During Use

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.

Recalled Product

  • Trevo XP ProVue Retriever
  • Catalog Numbers and Lot Numbers: Full List of Affected Devices
  • Manufacturing Dates: July 22, 2019 to September 9, 2019
  • Distribution Dates: September 18, 2019 to July 7, 2020
  • Devices Recalled in the U.S.: 1,258
  • Date Initiated by Firm: September 21, 2020

Device Use

The Trevo XP ProVue Retriever is intended to restore blood flow or remove blood clots within a blood vessel in the brain during an acute ischemic stroke.

Reason for Recall

Stryker Neurovascular is recalling the Trevo XP ProVue Retriever device because there is a risk the core wire may break or separate when the core wire is retracted during use. If this occurs, the device could be left inside the patient's blood vessel or tissue. The use of affected devices may cause serious adverse health consequences, including bleeding, additional blockage of blood vessels, disability, and death.

There have been 11 reports of injury or illness related to this device issue and one death.

Who May be Affected

  • Health care providers using the Trevo XP ProVue Retriever
  • Patients undergoing procedures involving the Trevo XP ProVue Retriever

What to Do

On September 21, 2020, Stryker Neurovascular sent a notice to all affected customers notifying them of the issue and provided the following instructions:

  • Immediately check your internal inventory for affected devices.
  • Segregate the affected units in a secure location for return to Stryker.
  • Circulate this Recall-Removal notice internally to all interested and affected parties.
  • Maintain awareness of this communication internally until all required actions have been completed within your facility.
  • Inform Stryker if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that Stryker can inform the recipients appropriately.
  • Inform Stryker of any adverse events concerning the use of the recalled devices.
  • Complete Stryker's customer response form and email it to NVCustomerCare@Stryker.com. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter.
  • Product return information will be provided to you by your designated Stryker territory manager or sales representative.

Contact Information

Customers with questions may contact Stryker Customer Care by email at NVCustomerCare@Stryker.com or by phone at 1-855-916-3876.

Additional Resources

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program using an online form, regular mail, or FAX.

Back to Top