U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Medical Device Recalls
  5. All Ultrasound Gels and Lotions Manufactured by Eco-Med Pharmaceutical, Inc. Recalled Due to Risk of Bacteria Contamination
  1. Medical Device Recalls

All Ultrasound Gels and Lotions Manufactured by Eco-Med Pharmaceutical, Inc. Recalled Due to Risk of Bacteria Contamination

The recall described in this notice is the same one that was announced in the Letter to Health Care Providers

11/23: This recall classification notice was expanded to include additional recalls. See updated list under Additional Resources.

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.

Recalled Product

  • Product Names: ultrasound gel products and lotions manufactured by Eco-Med are marketed by a number of distributors, under different product names, which include but are not limited to the following:
Product Name Distributors
Action de Gala FuSion Conductive Gel Aesthetics Systems (United States)
Athena Liaison Ultrasound Gel Mac Medical Supply Co Inc. (United States)
Conductor Transmission Gel Chattanooga Medical Supply (United States)
Conductor Australia DJO Australia (Australia)
DJO Conductor DJO Global (United States)
EcoGel 100 Amtec Sales, Inc. (dba Print Media) (United States)
Eco-Med Pharmaceutical (Canada)
Cardinal Health (Canada)
Fujifilm Sonosite, Inc. (United States)
Maranda Lauzon Inc (Canada)
Medline Canada (Canada)
NDC Inc. (United States)
STAT Healthcare Corporation (Canada)
The Stevens Company Limited (Canada)
EcoGel 200 Accelerated Care Plus (United States)
Active Crystal Inc (Canada)
Agencia Matamoros (Honduras)
Amesys S.A. (Honduras)
Amtec Sales, Inc. (dba Print Media) (United States)
Anamed Corp. (United States)
Best Medical Supply (United States)
BP Medical (Dominican Republic)
Cardinal Health (Canada)
Cardinal Health, Inc. (United States)
Choice Medical Systems, Inc. (United States)
Christie Innomed Inc. (Canada)
Clinomed, Inc. (United States)
CMCC Supply Centre and Bookstore (Canada)
Dada Dada & CIA (El Salvador)
Dectro International (Canada)
Dental & Medical Distributor (United States)
DHG Specialties (United States)
Dimesan USA (United States)
Dolphin Med-Equipment & Supply (United States)
Droredime (Honduras)
Dunbar Medical (Canada)
Dynamic Medical Supplies, Inc. (United States)
Eastern Medical Supply Ltd (Canada)
Eco-Med Pharmaceutical (Canada)
Equimedic S.A. de C.V. (El Salvador)
Eshus Merin, Inc (United States)
Euro Essential Ltd. (Canada)
Fernanada’s Beauty Products  (Canada)
Gadelius Medical (Japan)
Global Medical Solution (NZ) LTD (New Zealand)
Good-link Electronics Ltd. (Hong Kong)
Grobartig Solution, LLC. (United States)
Health & Global Enterprises (United States)
Healthy You (United States)
Henry Schein Arcona (Canada)
Innovaciones Medicas (Guatemala)
Intrasound Products, LLC (United States)
Inversiones C.A. (Honduras)
Luizapata Corp. (United States)
Lumenis Ltd (Israel)
Maranda Lauzon Inc (Canada)
Medical Depot, S.A. (Panama)
Medical Services, LLC. (United States)
Medical Specialists Co., Inc. (United States)
Medis Part Ltd. (Thailand)
Medline Canada (Canada)
Medplus Supply S.A. (Panama)
MJM Distributing (Canada)
Multiservicios Electromedicos, S.A. (Costa Rica)
NDC Inc. (United States)
North Coast Medical (United States)
Nova Skin, Inc. (United States)
Ortho Canada (Canada)
Owens & Minor Distribution, Inc. (United States)
Paramedicos del El Salvador (El Salvador)
PTmart, Inc. (United States)
Quantum Products, Inc. (United States)
RAF, S.A. DE C.V. (El Salvador)
Reserma, SA (Panama)
So medico SDN BHD (Malaysia)
Spavaro Inc. (Canada)
STAT Healthcare Corporation (Canada)
The Stevens Company Limited (Canada)
Trimedic Supply Network Ltd. (Canada)
Vitality Depot (Canada)
Westside Surgical Corporation (United States)
EcoGel 300 Eco-Med Pharmaceutical (Canada) 
MJM Distributing (Canada) 
Vitality Depot (Canada) 
EcoLotion Amtec Sales, Inc. (dba Print Media) (United States) 
Clinomed, Inc. (United States)
Eco-Med Pharmaceutical (Canada) 
Eshus Merin, Inc (United States)
Grobartig Solution, LLC. (United States)
Healthy You (United States)
Medical Services, LLC. (United States)
Medical Specialists Co., Inc. (United States)
NDC Inc. (United States)
North Coast Medical (United States) 
PTmart, Inc. (United States)
Quantum Products, Inc. (United States)
The Stevens Company Limited (Canada) 
Vitality Depot (Canada)
Intelect DJO Global (United States)
Liquasonic  Athena Medical Products Inc. (United States)
Cardinal Health Inc. (United States)
Concordance Healthcare Solutions (United States)
Henry Shein Inc. (United States)
Mckesson Medical Surgical Inc. (United States)
Medline Industries, Inc. (United States)
MAC Brand EcoGel  Mac Medical Supply Co Inc. (United States)
Medelco Medelco Inc. (Canada)
MediChoice Ultrasound Gel Mac Medical Supply Co Inc. (United States) 
North Coast Medical (United States)
Owens & Minor (United States)
Medline Ultrasound Gel Medline Industries Inc. (United States)
Myossage DJO Global (United States) 
North Coast Medical (United States)
Norco Ultrasound Gel
North Coast Ultrasound Gel
North Coast Medical (United States)
Omnisound Ultrasound Gel  Accelerated Care Plus (United States)
Pro Advantage Ultrasound Gel NDC Inc. (United States)
Red Medical Ultrasound Gel Red Medical Supplies Ltd. (Canada)
Scrip Crème All Purpose Lotion Scrip Hessco (United States)
Scrip Ultrasound Gel Scrip Hessco (United States)
Smart (EcoGel 200 Ultrasound Gel) Smart Technology and Product Co., Ltd (Thailand)
Ultra-Myossage DJO Global (United States)
  • Product Codes: MUI, ITX, GYB
  • Manufacturing Years: 2018 until Present
  • Distribution Years: 2018 until Present
  • Date Initiated by Manufacturer: August 4, 2021

Eco-Med has shut down all operations and is no longer manufacturing or distributing any products.

Device Use

Eco-Med ultrasound gels and lotions are used during an ultrasound test to create images of organs and structures inside the body. Ultrasound gel is a clear, water-based gel that is applied to the skin on the area to be examined. The gel helps with the transmission of sound waves.

Reason for Recall

All ultrasound gels and lotions manufactured by Eco-Med are being recalled due to risk of bacterial contamination with Burkholderia cepacia complex (Bcc). The FDA independently confirmed that distributed product tested positive for bacterial contamination. The use of affected ultrasound gels and lotions contaminated with Bcc may lead to serious infections, including bloodstream infections, which may result in sepsis or death.

As of Aug. 31, 2021, there have been at least 66 infections, including 60 bloodstream infections associated with these affected products, per the Centers for Disease Control and Prevention (CDC).

Who May be Affected

  • Health care providers who use ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc., and distributed by multiple entities under various product names.
  • Patients who require care using ultrasound gels or lotions manufactured by Eco-Med Pharmaceutical, Inc., and distributed by multiple entities under various product names.

What to Do

On Aug. 4, 2021, Eco-Med Pharmaceutical, Inc. sent a recall classification notice for certain lots of EcoGel, MediChoice, and Mac Medical ultrasound gel. However, the FDA has determined that all ultrasound gels and lotions manufactured by Eco-Med are at risk for bacterial contamination.

The FDA recommends health care providers and facilities:

  • Immediately stop using and discard all ultrasound gel and lotion products manufactured by Eco-Med.
  • Do not purchase ultrasound gels or lotions manufactured by Eco-Med.
  • Consult Centers for Disease Control and Prevention (CDC) and professional society guidelines regarding use of ultrasound gel and appropriate ultrasound cleaning procedures (see Additional Resources).
  • Contact the distributor that supplied these products to you if you have questions about the disposal of the products.
  • Report any adverse reactions or quality problems experienced with the use of these products to the FDA's MedWatch Adverse Event Reporting program.

Contact Information

Customers in the U.S. with questions about this recall should contact the distributor who supplied the affected products.

Eco-Med has shut down all operations and is no longer manufacturing or distributing any products.

Full List of Affected Devices

Please see Medical Device Recall Database Entry below.

Additional Resources:

  1. Stop Using All Eco-Med Ultrasound Gels and Lotions Due to Risk of Bacterial Contamination – Letter to Health Care Providers
  2. Eco-Med Pharmaceutical, Inc. Recall Letter
  3. Eco-Med Pharmaceutical, Inc. Recall Web Page
  4. Related FDA recall classification notice
  5. Accelerated Care Plus Corporation Recall Database Entry
  6. AMTEC Sales Inc. Recall Database Entry
  7. Aesthetics Systems USA, Inc. Recall Database Entry
  8. Athena Medical Products Inc. Recall Database Entry
  9. DJO, LLC Recall Database Entry
  10. Lumenis, Inc. Recall Database Entry
  11. MAC Medical Supply Co. Inc. Recall Database Entry
  12. National Distribution & Contracting, Inc. Recall Database Entry
  13. North Coast Medical Inc. Recall Database Entry
  14. Red Medical Supplies Recall Database Entry
  15. Scrip Inc. Recall Database Entry

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program using an online form, regular mail, or FAX.

 
Back to Top